Dissociation of solid tumors, like CT26 colorectal carcinoma, enables the tumor microenvironment to be explored at the single-cell level, providing valuable insights into immune response and disease progression. Accurate analysis of solid tumors requires efficient and effective dissociation to maintain the integrity and heterogeneity of the native cell population. Efficiency and efficacy are improved in an acoustic dissociation on the SimpleFlow, as compared to an enzymatic digestion.
The choice of dissociation method is crucial in determining the quality of the resulting single-cell suspension (1). Enzymes are known to bias solid tumor dissociations through selectively releasing certain types of cells from tissue, inducing transcription with heat, and damaging cell surface epitopes (2). Flow cytometry requires the preservation of cell surface proteins. When these markers are stripped from cells, or whole populations are eliminated, cancer research is greatly inhibited by the biased single-cell suspension.
In this experiment, we demonstrate that a six-minute acoustic dissociation on the SimpleFlow better preserved CD45, CD4, CD19, CD27, CD100, CD229 and CD44 cell surface markers of CT26 cells. Tumors dissociated without enzymes on the SimpleFlow had up to six times the expression as measured by MFI fold change, compared to even a mild (2) enzymatic digestion with Collagenase IV and DNase.
The acoustic dissociation protocol presented here provides an alternative to enzymatic dissociation with comparable sample preparation and post-dissociation steps but eliminates the long incubations required to activate enzymes. This reduces the total workflow down from over an hour to 25 minutes as displayed in Figure 10. This fast, yet gentle dissociation shortens the time from dissociation to downstream analysis, thus reducing the overall stress on cells.
Figure 10. Comparison of the 25-minute SimpleFlow workflow to the 70-minute enzyme workflow.
To investigate the differences in single-cell quality between SimpleFlow’s acoustic dissociation and a gentle enzymatic dissociation, 5 × 105 CT26 cells were implanted subcutaneously into BALB/c mice and allowed to grow for approximately four weeks. Tumors were divided into ~100 mg samples and assigned to either the SimpleFlow acoustic or the Collagenase IV + DNase enzymatic dissociation method. For the SimpleFlow acoustic dissociation, replicates were minced 200 times within the Cellsonics Dissociation Cartridge then subjected to low-amplitude acoustic pulsing while held in a temperature controlled 4 °C water bath for six minutes. For the enzymatic digestion samples were minced using a razor blade until homogenous, then incubated with Collagenase IV and DNase for 45 minutes at room temperature on the GentleMACS using a program with gentle mixing. Both groups were filtered through a 70-micron cell strainer and washed before proceeding with staining and flow cytometry on the BD Symphony. Data was processed using FlowJo Version 10.10.0 and data was analyzed using an unpaired two-tailed student’s t-test on GraphPad Prism 10.
Flow cytometry characterized many markers with reduced or altered expression when comparing the enzymatic dissociation to acoustic dissociation on the SimpleFlow, as described in Table 1. Due to the harsh enzyme dissociation, in some cases whole populations were missing or greatly reduced. For many of the events that maintained their positivity for certain markers, lower MFI signal was observed, indicating that many cell surface markers were damaged or removed from live cells when using an enzyme. Cleaved cell surface markers are evidence of poor cell health and have negative consequences for downstream analyses and applications like flow cytometry, flow activated cell sorting, CITE-seq, and outgrowth.
Even compared to an optimized, mild enzyme dissociation, SimpleFlow enhances the performance of flow cytometry as cell surface epitopes are better preserved with higher MFIs. Improved parenchymal representation and brighter signals push your cancer research forward by enabling faster and more dependable antibody-based analysis.
References